Ivoclar Programat P200: Manual
If you work with high-translucency zirconia or lithium disilicate, you’ve likely seen the Ivoclar Programat P200 . It’s a compact, high-temperature furnace designed for speed and precision. But here’s the catch: assuming it works like a standard porcelain furnace can ruin a case.
Deep Dive: The Ivoclar Programat P200 – A Manual Worth Reading (Even for Pros) Ivoclar Programat P200 Manual
| Code | Meaning | Fix (from manual troubleshooting) | |------|---------|-----------------------------------| | E-001 | Temp ramp too slow | Check elements & power supply | | E-017 | Vacuum not reached | Check door seal & pump oil | | E-105 | Cooling fan failure | Clean fan filter (rear panel) | | E-208 | Thermocouple open circuit | Replace thermocouple | If you work with high-translucency zirconia or lithium
| Material | Program name | Key manual note | |----------|--------------|----------------| | IPS e.max CAD crystallisation | e.max CAD | Use slow cooling for full translucency | | IPS e.max Press (invested) | e.max Press | Do use vacuum during pressing | | Zirconia (standard) | Zirlux 16 | Ramp rate 8–12°C/min – slower for >20mm spans | | Zirconia (high translucency) | Zirlux HT | Lower final temp by 10–15°C vs standard | Deep Dive: The Ivoclar Programat P200 – A
Always add a 5-minute drying hold at 300°C for thick zirconia (>2.5mm) to avoid microcracking. 5. The Maintenance Schedule (What Actually Wears Out)
Drop a comment or DM us. Disclaimer: This post is a technician’s summary. Always refer to the official Ivoclar Programat P200 manual for exact procedures and safety information.